Please help us provide you better service by rating this answer. Hardy Tech Support is always looking to improve service to our customers.

We can have someone at your facility when you need us! and Canada for system installation, startup, inspection, verification, calibration and certification, emergency troubleshooting, on-site training and routine preventative maintenance for your Process Weighing instrumentation (even for non-Hardy equipment). We offer Local Field Service in the continental U.S. If you need on-site assistance call Hardy Technical Support at 80 option #4 or 85 option #4. If you need additional information please click on the "Ask A Question" tab using Hardy's online WebTech knowledgebase.
ZERO INSTALL REMOTE PC CONTROL UPGRADE
The remote zero option can be ordered with a new controller with a special designation -119 or a field upgrade as the HI2151/30xx-119 If the TAG2=40 is 00 or another number contact Hardy technical support to verify the option was correctly ordered or if an ordering error has occurred. The second tag number (40) represents the remote zero option is enable. Enter the self test menu and look for the TAG1 and TAG2 entries. If the remote zero still does not function. Increase the auto zero setting or re calibrate the scale establishing a new gross zero. If auto zero function also does not work, the zero shift from the last calibration is greater than the auto zero tolerance setting. If remote zero does not function, enable auto zero tracking and see if that works. Only the upper left hand zero button uses the zero tolerance parameter. Auto zero tracking, communications zero command, and remote zero all use the auto zero parameter tolerance. The remote zero and auto zero tracking would fail to zero the additional 2 pounds on the scale. If you zero out 4 pounds gross using the zero button on the front display, and had the auto zero tolerance set for 2 lb. The NET weight has no limit and is set to zero using the tare function.Īs an Example: With a zero tolerance setting of 5 lb. The amount of zero is limited by the auto zero parameter and that zero limit is an accumulated amount of GROSS weight offset from the last calibration. The remote zero option is active with contact between the DIS terminal and ground terminal of the remote function terminal strip. Apply power and wait for the boot cycle to complete. Power down the controller, remover the old SMM and install the new SMM. With all the original setup parameters recorded.
ZERO INSTALL REMOTE PC CONTROL HOW TO
There should have been a addendum included that says how to install the SMM with power OFF. Installing the SMM under power will erase the TAG2 option code. You will need to re-enter all this information and re-calibrate the scale after the upgrade. Record any communications information and all the parameters from the old SMM using the front display to access the information.
ZERO INSTALL REMOTE PC CONTROL DRIVERS
Make sure your PC is updated with all the recent drivers and security updates. Go through our System Specification page and make sure your PC is qualified for your needs. Setup your system Step 1: System Requirements 1.

When installing a new Secure Memory Module (SMM) the instrument should treated as a new installation. Zero client installation and configuration 3.